- 111 -
Example:
Command:
OMI <CR><LF>
- return accessible working
modes
Response:
OMI <CR><LF>
19_”
Drying” <CR><LF>
- return accessible working
modes,
mode number + name
OK <CR><LF>
- command carried out
Set working mode
Command overview:
Command sets particular working mode.
Format: OMS_n <CR><LF>
Response options:
OMS_OK<CR><LF>
- command carried out
OMS_E <CR><LF>
- error in-course of command execution, no
parameter or incorrect format
OMS_I <CR><LF>
- command understood but not accessible at this
moment
n - parameter, decimal value determining working mode number. To see detailed description
go to OMI command.
Example:
Command:
OMS_19<CR><LF>
- set DRYING mode
Response:
OMS_OK<CR><LF>
- DRYING mode selected
Give current working mode
Command overview:
Command returns accessible working modes.
Format: OMG <CR><LF>
Response options:
OMG_n_OK <CR><LF>
-
command carried out, response: current working mode
OMG_I <CR><LF>
- command understood but not accessible at this moment
n - parameter, decimal value determining working mode number. To see detailed description
go to OMI command.
Example:
Command:
OMG<CR><LF>
- return current working mode
Response: OMG_19_”Drying”OK<CR><LF>
- moisture analyzer set to DRYING
mode
Give accessible units(weighing mode exclusively)
Command overview:
Command returns units available for a particular device and for a current working mode.
Format: UI <CR><LF>
Response options:
UI_”x
1
,x
2
, … x
n
”_OK<CR><LF>
- command carried out, returns accessible units
UI_I <CR><LF>
- command understood but not accessible at this moment
X - unit symbols, separated by means of commas