- value YES --,-- - (,) - comma is used as decimal point while
sending the mass value
- value NO - compatibility with other programs (mass value is
sent as a text,
DELAY if the transferred information is incomplete, i.e. part of it gets lost (due to
auto-entering or auto-formatting of data by computer program) during integration
with Excel type program, set the parameter value to <2>. If the transferred
information is still incomplete, the parameter value must be higher. The value
ranges between 0 (quick data transfer) and 9 (slow data transfer around 10
characters per second).
In case of Free Link printout (keyboard emulator) all languages without special
characters are operated (‘QWERTY’ keyboard layout), These languages do not use
special characters.
Additionally, the following languages are operated:
Polish for ‘Polish (Programmers)' keyboard. To enable support of this language
go to printer settings on your moisture analyzer and select Windows-1250 code
page (Central European, Latin-2). The following Polish diacritical signs are
added: ą, ć, ę, ł, ń, ó, ś, ż, ź.
German for ‘German (Germany)' keyboard. To enable support of this language,
go to printer settings on your moisture analyzer and select Windows -1252
code page (Western European, Latin-1). The following German diacritical signs
are added: ä, ö, ü, ß. The keyboard layout changes from 'QWERTY to
‘QWERTZ’. The characteristic feature of such keyboard is ‘ß’ sign located on
the right side of the ‘zero’.
Caution!
Free Link printout option allows printing only signs that are available on a keyboard.
Entering capital letters and special signs, by means of key combination where "Shift" key is
pressed, is also possible.
In case of Polish and German, basic special characters (written using key combination of a
given letter and right ‘Alt’ button) were added.
Free Link printout option does not support printing of remaining special signs.
An exemplary measurement printout is described in the PRINTOUTS section.
Additionally you can send a controlling code (of a hexadecimal form) to a printer either at the
beginning of the printout - <PREFIX> parameter, or at the end of it - <SUFFIX> parameter.
Sending these codes allows to control globally information and actions carried out at the
beginning and/or at the end of each printout sent from a weighing device to a printer.
This function is most frequently used:
- to send information about the code page of a printout sent by a moisture analyzer, at the
beginning, and
- to send a command enabling the crop of a paper in EPSON printers (if the printer is
equipped with an auto cutter blade), at the end.
Settings of <Prefix> and <Suffix> parameters are valid for all printouts sent via the weighing
device, e.g.: adjustment reports, statistics etc., and header, footer and GLP printouts.
Caution:
It must be remembered that inserting paper crop command to <SUFFIX> parameter (control
code) results in sending the code after each printout. If you wish for one whole printout to
consist of: HEADER, GLP PRINTOUT and FOOTER and to be cropped underneath the
FOOTER, then the paper crop command should be inserted only for FOOTER settings as a

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG MA X2.IC.A Series and is the answer not in the manual?

RADWAG MA X2.IC.A Series Specifications

General IconGeneral
BrandRADWAG
ModelMA X2.IC.A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals