non-standard printout with <{150}> variable (paper crop for EPSON printer). In such a case
<SUFFIX> command must remain empty.
To ensure correct integration of the moisture analyzer with the printer (correct printout of
diacritical signs of a given language), respective baud rate, obligatory for a given printer,
must be selected (read the printer settings). Additionally code page of a sent printout must
be accordant with a code page of a printer.
There are two methods for obtaining code page accordance:
by setting the right code page in the settings of a printer (read printer's user manual)
it must be accordant with the printout code page of a moisture analyzer:
Code page
Language
1250
POLISH, CZECH, HUNGARIAN
1251 or 866
RUSSIAN
1252
ENGLISH, GERMAN, FRENCH, SPANISH, ITALIAN
1254
TURKISH
1256
ARABIC
by sending the control code from the weighing device, which, prior the printout,
automatically sets the right code page of the printer, i.e. code page accordant with the
one of a weighing device (Read user manual of the printer you operate to check if it
supports this option).
CAUTION: CODES MUST BE ENTERED IN A HEXADECIMAL FORM!
Exemplary moisture analyzer settings for correct integration (printout of Polish signs) with
EPSON thermal printer connected to RS232 port:
1. EPSON TM-U220D
Communication parameters for port to which the printer is connected:
baud rate 9600 bit/s
parity none
Printer parameters set for PERIPHERALS group:
port COM 1 or COM 2 (the one to which the printer is connected)
code page 852
2. EPSON TM-T20
Communication parameters for port to which the printer is connected:
baud rate 38400 bit/s
parity none
Printer parameters set for PERIPHERALS group:
port COM 1 or COM 2 (the one to which the printer is connected)
code page 1250
prefix 1B742D
Printer’s parameters:
port COM 1 or COM 2 (the one to which the printer is connected)
code page 852
prefix 1B7412
If on the printout in the place of the last digit there are any unexpected signs (for verified
moisture analyzers), then <CONTROL CODES> parameter must incorporate, apart from