non-standard printout with <{150}> variable (paper crop for EPSON printer). In such a case
<SUFFIX> command must remain empty.
To ensure correct integration of the moisture analyzer with the printer (correct printout of
diacritical signs of a given language), respective baud rate, obligatory for a given printer,
must be selected (read the printer settings). Additionally code page of a sent printout must
be accordant with a code page of a printer.
There are two methods for obtaining code page accordance:
• by setting the right code page in the settings of a printer (read printer's user manual) –
it must be accordant with the printout code page of a moisture analyzer:
ENGLISH, GERMAN, FRENCH, SPANISH, ITALIAN
• by sending the control code from the weighing device, which, prior the printout,
automatically sets the right code page of the printer, i.e. code page accordant with the
one of a weighing device (Read user manual of the printer you operate to check if it
supports this option).
CAUTION: CODES MUST BE ENTERED IN A HEXADECIMAL FORM!
Exemplary moisture analyzer settings for correct integration (printout of Polish signs) with
EPSON thermal printer connected to RS232 port:
1. EPSON TM-U220D
Communication parameters for port to which the printer is connected:
• baud rate – 9600 bit/s
• parity – none
Printer parameters set for PERIPHERALS group:
• port – COM 1 or COM 2 (the one to which the printer is connected)
• code page – 852
2. EPSON TM-T20
Communication parameters for port to which the printer is connected:
• baud rate – 38400 bit/s
• parity – none
Printer parameters set for PERIPHERALS group:
• port – COM 1 or COM 2 (the one to which the printer is connected)
• code page – 1250
• prefix – 1B742D
Printer’s parameters:
• port – COM 1 or COM 2 (the one to which the printer is connected)
• code page – 852
• prefix – 1B7412
If on the printout in the place of the last digit there are any unexpected signs (for verified
moisture analyzers), then <CONTROL CODES> parameter must incorporate, apart from