EasyManua.ls Logo

R&S FPC Series - Page 271

R&S FPC Series
419 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
FPC
268User Manual 1178.4130.02 ─ 07
If you are using the average trace mode, you can set the number of measurements
with [SENSe:]SWEep:COUNt. Note that synchronization to the end of the average
count is possible only in single sweep mode.
Suffix:
<n>
.
irrelevant
<t> Trace
Parameters:
<TraceMode> You can turn off the trace with DISPlay<n>[:WINDow]:
TRACe<t>[:STATe] (trace mode "Blank").
For more information about the trace modes, see"Selecting the
trace mode" on page 116.
AVERage
"Average" trace mode.
BLANk
"Blank" trace mode.
MAXHold
"Max Hold" trace mode.
MINHold
"Min Hold" trace mode.
VIEW
"View" trace mode.
WRITe
"Clear / Write" trace mode.
*RST:
WRITe
Example: //Select trace mode
DISP:TRAC:MODE AVER
Manual operation: See "Trace Mode" on page 373
See "Trace Mode" on page 383
See "Trace Mode" on page 401
DISPlay<n>[:WINDow]:TRACe<t>[:STATe] <State>
This command turns a trace on and off.
Suffix:
<n>
.
irrelevant
<t> Trace
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF | 0
Example: //Turn on trace 2
DISP:TRAC2 ON
Remote Commands of the Spectrum Application

Table of Contents

Related product manuals