Remote Control - Commands
R&S
®
NRP2
230User Manual 1173.9157.02 ─ 03
CALCulate<[1]..8>:MAXimum:DATA?
CALCulate<[1]..8>:MINimum:DATA?
Queries the maximum/minimum of all measured values that previously occurred in this
calculate block.
The limit value is set to the current measured value when
●
the device is switched on
●
a reset is performed (*RST)
●
the value is explicitly set to the current measured value using the command
CALCulate[1..8]:EXTRemes:RESet.
The unit is the current unit of the measured result.
Possible Units: W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:
MEAN:DATA?
Queries the mean value of all measured values of the calculate block. The mean value
is reset when the auxiliary values are reset. The following formula applies:
Fig. 6-5: Mean value of all measured values
The unit is the current unit of the measured result.
Possible Units: W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:PTPeak:DATA?
Queries the peak-to-peak distance (maximum to minimum) of the measured values of
the calculate block.
The unit is the current unit of the measured result.
Possible Units: W | DBM | DBUV | DPCT | DB | O
Usage:
Query only
CALCulate<[1]..8>:RELative[:MAGNitude] <value>
Determines a value that is used as a divisor (logarithmic subtraction) for all measured
values of the calculate block in the CALC:REL:STAT ON state. The unit of this value is
equal to the unit of the output value of the calculate block and is defined according to the
following table:
CALCulate (Configuration of Calculate Blocks)