Remote Control - Commands
R&S
®
NRP2
266User Manual 1173.9157.02 ─ 03
[SENSe<[1]..4>]:CORRection:OFFSet <offset_value>
A fixed offset value can be defined for multiplying (logarithmically adding) the measured
value of a sensor. If the parameters are specified without a unit, the unit set by
UNIT:POW:RAT will be used.
Unit: DB | PCT
Parameters:
<offset_value> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: DB
[
SENSe<[1]..4>]:CORRection:OFFSet:STATe <state>
Switches the offset correction on or off.
Parameters:
<state> ON | OFF
*RST: depending on sensor
[SENSe<[1]..4>]:CORRection:DCYCle[:INPut][:MAGNitude] <magnitude>
Informs the R&S NRP2 about the duty cycle of the power to be measured. The duty cycle
correction is only applied in the ContAv mode where measurements are performed con-
tinuously without taking the timing of the signal into account.
Parameters:
<magnitude> float_value
Range: depending on sensor
*RST: depending on sensor
Default unit: PCT
[SENSe<[1]..4>]:CORRection:DCYCle:STATe <state>
Switches duty cycle correction for the measured-value on or off.
Parameters:
<state> ON | OFF
*RST: depending on sensor
SENSe (Sensor Commands)