Remote Control - Commands
R&S
®
NRP2
316User Manual 1173.9157.02 ─ 03
SYSTem:BEEPer:NOTify:OVERload <state>
Determines if the R&S NRP2 shall emit an acoustic signal when the power at a sensor
exceeds the power limit.
Parameters:
<state> ON | OFF
*RST: OFF
Usage: Event
SYSTem:
BEEPer:NOTify:KEY <state>
Enables (ON) or disables (OFF) acoustic key clicks, respectively.
Parameters:
<state> ON | OFF
*RST: OFF
SYSTem:BEEPer:STOP
Stops the output of an audible signal.
Usage:
Event
SYSTem:BEEPer:TIME <nrf>
Sets the length in milliseconds of the audible signal output with SYS:BEEP.
Parameters:
<nrf> int_value
Range: 1...60
*RST: This setting is not changed by *RST.
Default unit: S
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <nr1>
Sets the address with which the R&S NRP2 can be addressed via the IEC/IEEE bus. The
address is factory-set to 20 and is not changed by a reset.
Parameters:
<nr1> int_value
Range: 0...30
*RST: This setting is not changed by *RST.
Usage: Event
SYSTem