Remote Control - Commands
R&S
®
NRP2
318User Manual 1173.9157.02 ─ 03
SYSTem:COMMunicate:INET[:SELF]:SUBNetmask:ADDRess
<sm1>,<sm2>,<sm3>,<sm4>
Sets the subnet mask of your local subnet.
Note: This command cannot be used, while VXI11 or Ethernet is in use.
Note: Choosing an invalid subnet mask address may disturb the traffic on your LAN. If
you are not sure how to configure these settings, please ask your network administrator.
Parameters:
<sm1...4> int_value
Range: 0...255
*RST: This setting is not changed by *RST.
Usage: Event
SYSTem:
DATE <year>,<month>,<day>
Sets the date. Since the R&S NRP2 has no battery-backed clock, the date has to be reset
if the clock was not set after the last power-on of the R&S NRP2.
The date is set in the format yyyy,(m)m,(d)d. This setting is not changed by *RST.
Table 6-33: Parameter range
Parameter Value range
<year> 2000 ... 2100
<month> 1 (January) ... 12 (December)
<day> 1 ... 31
SYSTem:IDN <string>
Determines the user defined return value for the *IDN query. The maximum string length
is 128 characters.
SYSTem:IDN:AUTO <state>
Activates/deactivates return of the user-defined string for the *IDN query.
SYSTem:INFO? <string>
Returns information about the system. <string_value> is used to query a specific
information item. If called without parameters, the command returns all available infor-
mation in the form of a list of strings separated by commas. <string_value> can have
the following values. No difference is made between upper-case and lower-case letters:
SYSTem