Remote Control - Commands
R&S
®
NRP2
319User Manual 1173.9157.02 ─ 03
<string_value> Meaning
"Manufacturer"
Manufacturer of device
"Type"
Type (R&S NRP2)
"Stock Number"
Order No. of R&S NRP2
"Serial"
Serial number
"HWVersion"
Hardware version
"SW Build"
Firmware version
"BootloadVer"
Version of boot loader
"KeybCtrlVer"
Version of keyboard controller
"Options"
Installed options
"MACAddr"
MAC address of the R&S NRP2 if Ethernet option has been installed
"RAMSize"
RAM size
"ROMSize"
ROM size
"CPUClock"
Clock frequency of processor
"CPLDVersion"
Version of CPLD chip
"PICVersion"
Version of PIC-Chip
Parameters:
<string> string_value
Usage: Query only
SYSTem:ERRor?
This query returns error numbers and text from the error/event queue of the R&S NRP2.
A specific error number with text is entered in the error/event queue for each error that
occurs in the R&S NRP2. The queue functions according to the FIFO principle. The mes-
sage entered first in the queue is also output first when a SYST:ERR? query is sent.
Querying clears the message.
Up to 30 errors can be stored in the error/event queue. If further errors occur, the most
recent error is overwritten by the message -350, "Queue overflow".
If the error/event queue is empty, the message 0, "No error" is returned as a response.
The queue is cleared by *CLS. It is not changed by*RST.
For further information on the error/event queue see chapter 6.13, "STATus",
on page 292.
Usage:
Query only