Remote Control - Commands
R&S
®
NRP2
321User Manual 1173.9157.02 ─ 03
Simulated front-panel key <character_value> <NR1>
DOWN 22
LEFT 23
RIGHt 24
Note:
The R&S NRP2 can be switched off via remote control by means of the SYSTem:KEY
POWER command.
Parameters:
<key_stroke> <NR1> | <character_value> (see table)
<NR1>
System will be switched off SYSTem:KEY 17
<character_value>
System will be switched off SYSTem:KEY POWER
Usage: Event
SYSTem:LANGuage <emulation>
Selects an emulation of a different command set. The R&S NRP2 is able to emulate
different power meters by changing the remote command set. As the list of supported
instruments is extended continuously, the emulations are documented in separate appli-
cation notes.
Parameters:
<emulation> character_data
NRP | NRP2 | HP436A | HP437B | HP438A | E4418B | E4419B
| N1911A | N1912A
Usage: Setting only
SYSTem:OPT <string>
Determines the user defined return value for the *OPT query. The maximum string length
is 128 characters.
SYSTem:OPT:AUTO <state>
Activates/deactivates return of the user-defined string for the *OPT query.
SYSTem:PRESet
Resets the R&S NRP2 to default (*RST) values. No difference is made between reset
and preset values (exceptions: INIT:CONT, SENS:AVER:TCON, and
SENS:TRAC:AVER:TCON). For preset values see table 6-2.