Remote Control - Commands
R&S
®
NRP2
322User Manual 1173.9157.02 ─ 03
Usage: Event
SYSTem:REBoot
Reboots the R&S NRP2.
Usage: Event
SYSTem:SENSor<[1]..4>:INFO?
Returns information on a sensor. <string> is used to query a specific information item. If
called without parameters, the command returns all available information in the form of
a list of strings separated by commas. The permissible values for <string> depend on the
sensor. For details refer to the documentation of the corresponding sensor.
Parameters:
<string> Range: depending on sensor
Usage: Query only
SYSTem:SENSor<[1]..4>:REBoot
Performs a power down reset of the selected sensor.
Usage:
Event
SYSTem:SENSor<[1]..4>:RESet
From the point of view of the R&S NRP2 base unit, the sensors are stand-alone meas-
uring devices. They communicate with the R&S NRP2 via a command set complying with
SCPI.
SYST:SENS<[1].48>:RES prompts the base unit to send an *RST to the sensor in
question. Measurements in progress are interrupted.
Usage:
Event
SYSTem:SENSor<[1]..4>:TEST?
Performs a sensor test and returns a list of strings separated by commas. The contents
of this test protocol is sensor-specific. For its meaning, please refer to the sensor docu-
mentation.
Usage:
Query only
SYSTem:SPEed <state>
The data processing speed of the R&S NRP2 can be increased when FAST is selected.
The display is switched off and the measured values are no longer displayed since the
continuous update of the screen content requires computation time.
SYSTem