Press IP, and then the numeric keypad is displayed. Input the desired IP address
with the numeric keypad. This setting will be saved to the non-volatile memory; if
"Power On" is set to "Last", then DHCP and Auto IP are disabled at the next power-
on. The oscilloscope will load the preset IP address automatically.
2. Set the Subnet Mask
The format of the subnet mask is nnn.nnn.nnn.nnn. Wherein, the range of "nnn" is
from 0 to 255. You are recommended to ask your network administrator for a
subnet mask available.
Press
SubMask, and then the numeric keypad is displayed. Input the desired
subnet mask with the numeric keypad. This setting will be saved in the non-volatile
memory; if "Power On" is set to "Last", then DHCP and Auto IP are disabled at the
next power-on. The oscilloscope will load the preset subnet mask automatically.
3. Set the Gateway
You can can set this parameter in Static IP mode.
The format of the gateway is nnn.nnn.nnn.nnn. The range of the first segment
(nnn) is from 0 to 223 (except 127), and the range for the other three segments is
from 0 to 255. You are recommended to ask your network administrator for a gate
address available.
Press
Gateway, and then the numeric keypad is displayed. Input the desired
subnet mask with the numeric keypad. This setting will be saved in the non-volatile
memory; if "Power On" is set to "Last", then DHCP and Auto IP are disabled at the
next power-on. The instrument will load the preset gateway automatically.
4. Set the DNS
You can set this parameter in Static IP mode.
The format of the DNS address is "nnn.nnn.nnn.nnn". The range for the first
segment (nnn) of the address is from 0 to 223 (except 127); and the range for the
other three segments is from 0 to 255. You are recommended to ask your network
administrator for an address available.
Press
DNS, and then the numeric keypad is displayed. Input the desired DNS
address with the numeric keypad. Generally, you do not need to set the DNS,
therefore this parameter setting can be ignored.
TIP
When the three IP configuration types are all turned on, the priority of the parameter
configuration from high to low is "DHCP", "Auto IP", and "Static IP".
The three IP configuration types cannot be all turned off at the same time.
Apply the Network Parameter Setting
Press
Apply to validate the current network parameter setting.