When the measurement source is set to "LA", you need to set the LA display type.
Press More, and then press La Type continuously to select "HEX" or "BIN".
6. Measurement Example
Measure the period of a sine wave by using the manual cursor measurement and
Auto measurement respectively. The measurement results are both 1 ms, as shown
in
Figure 9.3
.
Auto Measurement
Cursor Measurement
Figure 9.3 Manual Cursor Measurement Example
9.3.2 Track Mode
In this mode, you can adjust the two pairs of cursors (Cursor A and Cursor B) to
measure the X and Y values on two different sources respectively. The points being
measured on Cursor A and Cursor B are marked by and respectively. When
the cursors are moved horizontally/vertically, the markers will position on the
waveform automatically. When the waveform is expanded or compressed
horizontally/vertically, the markers will track the points being marked at the last
adjustment of the cursors.
NOTE
When the delayed sweep function is enabled, the track cursor is within the expanded area.
Press Mode to select "Track" to enable the cursor track function. The measurement
results will be displayed at the upper-left corner of the screen, as shown in the
following figure. When you modify the Track measurement parameter, the
measurement results will be changed accordingly.
Operations and Measurements
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
193