Roche Diagnostics
E-14 Operators Manual · Version 1.0
9 Instrument alarms (messages) URISYS 2400
Instrument alarms (messages)
12 DC Power
DC Power Supply Error
1-2 STOP 1. Contact Technical Support.
3 E.STOP 1. Contact Technical Support.
13 Water Supply Tank
Water Sensor Error
1 STOP 1. Check that the cable is connected.
Switch off the power.
2. Attach the connector to the water
supply tank.
3. Switch on the power.
14 Short of water
No Water Error
1S.STOP
STOP
1. Check the float switch.
2. Add rinsing water to the water supply
tank, and try to restart.
2 WARNING 1. Check the float switch.
2. Add rinsing water to the water supply
tank, and try to restart.
15 Liquid Waste Tank
Waste Sensor Error
1 STOP 1. Check that the cable is connected.
2. Switch off the power.
3. Attach the connector to the liquid
waste tank.
4. Switch on the power.
16 Liquid Waste Tank
full
Full Waste Error
1 S.STOP 1. Empty the liquid waste tank.
17 Watch Dog Timer
Timing Error
1 E.STOP 1. Create a backup file of alarm trace data
on a new diskette, by executing the FD
write procedure on the [Alarm Trace]
screen.
2. Contact Technical Support.
19 TS Cassette empty
No Test Strip
1 S.STOP 1. Replace the URISYS 2400 cassette.
2 WARNING 1. Test Strip number below alarm level
3 WARNING 1. Test Strip number below alarm level
20 STAT Position
STAT Position Error
1 WARNING 1. Place the STAT sample at the STAT
sampling position, and request a STAT
measurement.
3 WARNING 1. Place the washing solution at the STAT
sampling position, and perform the
wash procedure.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)