Roche Diagnostics
Operator’s Manual · Version 1.0 E-13
URISYS 2400 9 Instrument alarms (messages)
Instrument alarms (messages)
3 Supply Line
Rack In Error
1-5, 7 S.STOP 1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
6 STOP 1. Place the rack for SG.
2. Execute SG calibration again.
4Crossfeed Line
Rack Crossfeed Line
Error
1-6 S.STOP 1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
5Stock Line
Rack Out Line Error
1-3 1. Execute the reset procedure. If the
condition cannot be restored to normal,
contact Technical Support.
4 STOP 1. Place the rack tray on the rack input
line.
5 S.STOP 1. Place the rack tray on the rack input
line.
6 Photometer
Photometer Error
1-2, 4 STOP 1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
7 TS Transfer
Test Strip Transport
Error
1-7 STOP 1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
8S.Syringe
Sample Syringe Error
1-2 STOP 1. Execute the reset procedure.
2. If the condition cannot be restored to
normal, contact Technical Support.
9 Waste Box
No Waste Box
1 WARNING 1. Insert the strip waste container into the
analyzer.
2. Restart.
2 STOP 1. Insert the strip waste container. Insert
the strip waste container into the
analyzer.
10 Waste Box full
Full Waste Box Error
1 WARNING 1. Empty the strip waste container.
2. Insert the strip waste container into the
analyzer.
2 S.STOP 1. Empty the strip waste container.
2. Insert the strip waste container into the
analyzer.
11 Fuse
Fuse Blown Error
1-3 E.STOP 1. Contact Technical Support.
Category Alarm name
Type of error
Subcategory Level Check and countermeasure
Table E-3 Instrument Alarms (messages) (continued)