Roche Diagnostics
Operator’s Manual · Version 1.0 B-31
URISYS 2400 4 Cassette
First-time installation
9 Open the cassette compartment door on the analyzer.
10 Place the cassette on the opened door and move it smoothly into the
cassette compartment with the label facing out.
11 Move the cassette completely to the rear of the cassette compartment,
ensure it fits correctly, and close the door firmly.
12 Press <Yes> on the window screen for confirmation to return to the
[Change Cassette] screen.
13 Enter the Lot Number of the cassette from the outer packaging box. Press
<Enter> to confirm.
14 Enter for “Alarm level” the number of remaining strips that will lead to an
alarm message indicating a low number of strips (default setting is 40).
15 Press <New> to set the cassette mechanism to the working position. A
window “Do not open the door” appears.
16 After completing the setting, a confirmation window “Initialization
complete, press <OK> to continue” appears.
17 Press <OK> to return to the [Change Cassette] screen.
Figure B-12 URISYS 2400 cassette with protective covers
Figure B-13 Inserting cassette into URISYS 2400 analyzer