Instrument Function
R&S
®
SMC100A
88Operating Manual 1411.4060.32 ─ 09
Display Update is On/Off
Switches on/off update of the displayed parameters.
Switching off the update of the displayed parameters increases the speed for certain
settings.
Note: For optimum sweep performance with short dwell times and for fast settling
times, it is recommended to switch off the display update .
Remote command:
:SYSTem:DISPlay:UPDate on page 359
5.2.3.7 Selftest
A selftest is provided for service purposes.
Access:
1. Select "Setup > Test > Selftest"
2. Select "Start Selftest".
The following tests are only available via remote control:
●
:TEST<hw>:DIRect on page 363
Start Selftest
Performs a selftest on all installed hardware options.
When completed, the R&S SMC displays a list of all performed test cases and the test
results (passed or failed).
Note: While the self test is in progress, the actual signal level at the RF output is -50
dBm. This value is not indicated in the status bar.
Remote command:
:TEST<hw>:ALL:STARt on page 362
:TEST<hw>:ALL:RESult? on page 362