Remote Control Commands
R&S
®
SMC100A
333Operating Manual 1411.4060.32 ─ 09
5. Select the sweep mode and activate the sweep.
[SOURce:]SWEep[:FREQuency]:MODE AUTO
[SOURce:]FREQuency:MODE SWEep
6. Trigger the sweep.
[SOURce:]SWEep[:FREQuency]:EXECute
It is recommended that you switch off the "Start/Stop Display Update" for optimum
sweep performance, especially with short dwell times (SYSTem:DISPlay:UPDate
OFF).
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl..................................................................333
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute.............................................................. 334
[:SOURce<hw>]:SWEep[:FREQuency]:MODE.................................................................. 334
[:SOURce<hw>]:SWEep[:FREQuency]:POINts................................................................. 335
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace..............................................................335
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?............................................................ 336
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe................................................................. 336
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing...............................................................336
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear].......................................................337
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic................................................ 338
[:SOURce<hw>]:SWEep:POWer:DWELl.......................................................................... 339
[:SOURce<hw>]:SWEep:POWer:EXECute....................................................................... 339
[:SOURce<hw>]:SWEep:POWer:MODE...........................................................................339
[:SOURce<hw>]:SWEep:POWer:POINts.......................................................................... 340
[:SOURce<hw>]:SWEep:POWer:RETRace.......................................................................341
[:SOURce<hw>]:SWEep:POWer:RUNNing?..................................................................... 341
[:SOURce<hw>]:SWEep:POWer:SHAPe..........................................................................341
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?........................................................... 342
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic]....................................................... 342
[:SOURce<hw>]:SWEep:RESet[:ALL].............................................................................. 343
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
Sets the time taken for each frequency step of the sweep.
The keyword [:FREQuency] can be omitted (see example). The command is then
SCPI-compliant.
Tip: It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (SYSTem:DISPlay:UPDate OFF).
Parameters:
<Dwell> float
Range: 2E-3 to 100
Increment: 100E-6
*RST: 15E-3
Example:
SWE:DWEL 12 ms
sets a dwell time of 12 ms for a frequency sweep at the RF out-
put.
SOURce Subsystem