Instrument Function
R&S
®
SMC100A
90Operating Manual 1411.4060.32 ─ 09
Check that the key in the dialog turns red.
Note: Pressing the same key again has no further effect, with the exception of the
ESC key, see Press the ESC key a third time.
5. Continue with the next key on the front panel and repeat step 3 to step 5 until all
keys are tested.
The test is completed, when each key is verified successfully, confirmed by a "Test
passed" message.
Select "OK" to exit the test.
Press the ESC key a third time.
Exits the "Check Front Panel" dialog, even if you have not yet checked all the keys.
Expected responses:
Pressing a key once (green), pressing twice (red)
Pressing the ESC key a third time exits the dialog.
If you detect a malfunction, for example, you press the front panel key the first time,
and the color of the button in the dialog turns red (instead of green), the front panel key
may be stuck. In this case, contact the Rohde & Schwarz Customer Support Center for
technical support, see http://www.customersupport@rohde-schwarz.com.
5.2.3.9 Shutting Down and Rebooting the Instrument
The POWER ON/STANDBY front panel key switches the instrument from the standby
to the ready state or vice versa. In remote operation form a remote computer or in
manual control, the R&S SMC provides you with another possibility to shut the instru-
ment down or to reboot the system.
To access the required settings, select "Setup > Environment > Shut Down".
Remote control commands:
:SYSTem:SHUTdown on page 359
:SYSTem:REBoot on page 358
see also :SYSTem:RESTart on page 358