Remote Control Commands
R&S
®
SMC100A
359Operating Manual 1411.4060.32 ─ 09
:SYSTem:SHUTdown
Shuts down the instrument.
Usage: Event
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete> 0 | 1 | OFF | ON
*RST: 0
Example:
SYST:STAR:COMP
Response: 1
The startup of the instrument is completed.
Usage: Query only
:SYSTem:DISPlay:UPDate <Update>
Switches the update of the display on/off. A switchover from remote control to manual
control always sets the status of the update of the display to ON.
Parameters:
<Update> 0 | 1 | OFF | ON
*RST: ON
Example:
SYST:DISP:UPD OFF
switches update of displayed parameter values off.
Manual operation: See "Display Update is On/Off" on page 88
:SYSTem:DATE <Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
This parameter is protected, in order to prevent accidental changes.
It can be accessed with protection level 1, see :SYSTem:PROTect<ch>[:STATe]
on page 358.
Parameters:
<Year> <year>,<month>,<day>
<Month> integer
Range: 1 to 12
<Day> integer
Range: 1 to 31