Remote Control Commands
R&S
®
SMC100A
358Operating Manual 1411.4060.32 ─ 09
Note: While working in a emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYSTem:LANGuage will be discarded.
The return to the SCPI command set of the R&S SMC can only be performed by using
the appropriate command of the selected command set. For example, the HP com-
mand EX returns to the instrument-specific GPIB command set (selection SYST:LANG
'HPxxxx').
Parameters:
<Language> string
Example:
SYSTem:LANGuage "SCPI"
sets the SCPI command set.
Manual operation: See "Language" on page 97
:SYSTem:PROTect<ch>[:STATe] <State>[, <Key>]
Activates/deactivates the specified protection level.
Parameters:
<State> select
*RST: 1
Setting parameters:
<Key> integer
The respective functions are disabled when the protection level
is activated. No password is required for activation. A password
must be entered to deactivate the protection level. The pass-
word for the first level is 123456.
This protection level can be used to lock-out internal adjust-
ments.
Example:
SYSTem:PROTect1:STATe ON
activates protection level 1. Internal adjustments are only possi-
ble after deactivating the lock-out.
SYSTem:PROTect1:STATe OFF,123456
deactivates protection level 1. Internal adjustments are enabled
again.
Manual operation: See "Protection Level/Password" on page 99
:SYSTem:REBoot
Restarts the firmware and the operating system.
Usage:
Event
:SYSTem:RESTart
Restarts the firmware. The operating system remains active.
Usage:
Event
SYSTem Subsystem