Instrument Function
R&S
®
SMC100A
97Operating Manual 1411.4060.32 ─ 09
You find all the remote control command sets supported by the R&S SMC in a selec-
tion list. For more information on this topic, an application note describes in detail how
to use this feature. See the product site of the R&S SMC.
The selected instrument also defines the identification string that is retrieved with query
*IDN?. In addition to the preset values, you can enter a user-defined identification
string, for example to provide individual identification for each generator, like
'MY_R&S SMC' (see Mode and IDN String).
As any other parameter, you can additionally change the remote control command set
to be emulated via the Language command. However, once you have switched to an
emulation, the R&S SMC specific command set is disabled, that means this command
is no longer effective. To return, you need to know the corresponding remote control
command of the simulated instrument. If you emulate an HP generator for example,
the HP command EX returns to the SCPI command set.
To access this dialog, press the SETUP or MENU key and select "Remote > Instru-
ment Emulations".
The "Instrument Emulations" dialog enables you to emulate a remote control com-
mand set of several other signal generators.
The remote commands required to remotely configure the emulation settings are
described in Chapter 7.15, "SYSTem Subsystem", on page 347.
Language
Selects the instrument whose remote command set is emulated by the R&S SMC.
Remote command:
:SYSTem:LANGuage on page 357
Mode
Selects the way the instrument identification is performed.
"Automatic"
Sets the "IDN String" and the "OPT String" automatically for the
instrument selected with the parameter Language.
"User Defined"
Enables you to define the "IDN String" and the "OPT String" for the
instrument selected with the parameter Language.
Remote command:
:SYSTem:IDENtification on page 357