Remote Control Commands
R&S
®
SMC100A
265Operating Manual 1411.4060.32 ─ 09
7.11 OUTPut Subsystem
The OUTPut system contains the commands which set the properties of the RF output
connector.
The properties of the LF output connector are set in the SOURce:LFOutput Subsystem
system.
:OUTPut<hw>:AFIXed:RANGe:LOWer?........................................................................... 265
:OUTPut<hw>:AFIXed:RANGe:UPPer?............................................................................265
:OUTPut<hw>:AMODe................................................................................................... 266
:OUTPut<hw>:IMPedance?.............................................................................................266
:OUTPut<hw>:PROTection:CLEar................................................................................... 266
:OUTPut<hw>:PROTection:TRIPped?..............................................................................267
:OUTPut<hw>[:STATe]...................................................................................................267
:OUTPut<hw>[:STATe]:PON........................................................................................... 267
:OUTPut<hw>:AFIXed:RANGe:LOWer?
Queries the minimum level which can be set when the attenuator is fixed, see :
OUTPut<hw>:AMODe.
Return values:
<Lower> float
Increment: 0.01
Example:
OUTP:AFIX:RANG:LOW?
queries the minimum level for the FIXed setting.
Example:
Response: -50
The minimum level is -50 dBm.
Usage: Query only
Manual operation: See " Fixed Range (PEP) In " on page 132
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set when the attenuator is fixed, see :
OUTPut<hw>:AMODe.
Return values:
<Upper> float
Increment: 0.01
Example:
OUTP:AFIX:RANG:UPP?
queries the maximum level for the FIXed setting for the RF out-
put.
Example:
Response: -27
The maximum level is -27 dBm.
Usage: Query only
OUTPut Subsystem