Rohde & Schwarz SMC100A - [:Source<Hw>]:Sweep[:Frequency]:Step[:Linear]

Rohde & Schwarz SMC100A
413 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMC100A
337Operating Manual 1411.4060.32 ─ 09
Parameters:
<Spacing> LINear | LOGarithmic
LINear
With the linear sweep, the step width is a fixed frequency value
which is added to the current frequency. The step width for lin-
ear sweep is entered in Hz (see [:SOURce<hw>]:SWEep[:
FREQuency]:STEP[:LINear] on page 337).
LOGarithmic
With the logarithmic sweep, the step width is a constant fraction
of the current frequency. This fraction is added to the current fre-
quency. The logarithmic step width is entered in % (see [:
SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
on page 338).
*RST:
LINear
Example:
SWE:SPAC LIN
selects linear sweep spacing for a frequency sweep at the RF
output.
Manual operation: See " Spacing - Frequency Sweep " on page 161
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step size for linear RF frequency sweep steps.
This parameter is related to the number of steps ([:SOURce<hw>]:SWEep[:
FREQuency]:POINts) within the sweep range as follows:
f
STARt
< f
STOP
freq_points = (f
SPAN
/ step_lin) + 1
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
The keywords [:FREQuency] and [:LINear] can be omitted. The command is then
SCPI-compliant.
Parameters:
<Linear> float
Range: full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting