Remote Control Commands
R&S
®
SMC100A
270Operating Manual 1411.4060.32 ─ 09
Example:
SENS:UNIT DBM
selects unit dBm for presentation of measurement result.
READ1?
queries the measurement result of the sensor connected to the
SENSOR interface.
Response: -45.6246576745440230
-45.6 dBm were measured at the given frequency.
or e.g. for R&S NRP-Z81
Response:
-55.62403263352178,-22.419472478812476
-55,6 dbm is the measured average level, -22. 4 dBm is the
measured peak level at the given frequency
Usage:
Query only
Manual operation: See " Level (Peak) " on page 151
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
<State>
The command activates the use of the s-parameters correction data of the selected
power sensor.
Note: For power sensor with attenuator this command is automatically set to ON.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: OFF
Example:
SENS:POW:CORR:SPD:STAT ON
activates the use of the s-parameters correction data of power
sensor 1.
Manual operation: See " Use SParameter - Power Sensors " on page 139
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe <State>
The command switches on and off the permanent indication of the power measure-
ment result in the upper right corner of the block diagram. For each sensor, the type of
sensor, the connector, the measurement source and - if set - the offset is indicated.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: ON
Example:
SENS1:POW:DISP:PERM:STAT ON
the permanent viewer is switched on.
Manual operation: See " Permanent Display State " on page 151
SENSe, READ and INITiate Subsystems