Instrument Function
R&S
®
SMC100A
133Operating Manual 1411.4060.32 ─ 09
Remote command:
[:SOURce<hw>]:POWer:STEP:MODE on page 322
Variation Step
Sets the user-defined step width for entering the RF level using the rotary knob. Level
variation with this step width must also be activated with "Variation Active".
Remote command:
[:SOURce<hw>]:POWer:STEP[:INCRement] on page 321
External Level Adjustment
The external level adjustment provides information about the data that has been used
for calibrating the RF level.
By default the instrument uses correction data obtained in the factory before delivery.
In exceptional cases, you can determine the calibration values with an R&S NRP
power sensor, and use these values for the external level correction. This feature is a
protected function (see Service Manual, chapter 2, "Adjustment").
Adjustment Data
Indicates what data has been used for level calibration.
Remote command:
:CALibration<hw>:LEVel:EXTern:DATA on page 241
5.3.5.3 Power-On/EMF Settings
The power-on behavior of the R&S SMC and the level display in the display header are
set in the "Power-On / EMF Settings" section of the "RF Level/EMF" dialog.
To open the "RF Level/EMF" dialog, select "RF > Configure > EMF" or use the MENU
key under "RF".
Power-On State - RF Signal
Selects the state which the RF output is to assume after the instrument is switched on.
"RF Off"
The output is deactivated when the instrument is switched on.
"Previous Set-
ting"
When the instrument is switched on, the output assumes the same
state as it had when the instrument was switched off.
Remote command:
:OUTPut<hw>[:STATe]:PON on page 267
Display Level as Voltage of EMF - RF Level
Activates display of the signal level as voltage of the EMF (no-load voltage). If this set-
ting is deactivated, the level is displayed as a voltage over a 50 Ohm load.
RF Block