EasyManua.ls Logo

Rohde & Schwarz SMC100A - MMEMory Subsystem

Rohde & Schwarz SMC100A
413 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMC100A
241Operating Manual 1411.4060.32 ─ 09
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Usage: Query only
Manual operation: See "Adjust Synthesis" on page 82
:CALibration<hw>:LEVel[:MEASure]? [<Force>]
Starts all adjustments which affect the level.
The acquired correction values improve the settling time and the signal quality.
Query parameters:
<Force> string
*RST: force
Return values:
<Measure> 0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage: Query only
Manual operation: See "Adjust Level" on page 82
:CALibration<hw>:LEVel:EXTern:DATA <Data>
Queries what data has been used for the level calibration.
By default the instrument uses correction data obtained in the factory before delivery.
In addition, customer data can be used for external level correction. The customer data
is obtained using a R&S NRP power sensor. External level correction is a protected
function (see service manual, chapter 2, "Adjustment").
Parameters:
<Data> FACTory | CUSTomer
*RST: FACTory
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external
level correction.
Manual operation: See "Adjustment Data" on page 133
:CALibration<hw>:ROSCillator[:DATA] <Data>
Sets the calibration value for the custom defined external adjustment.