Remote Control Commands
R&S
®
SMC100A
354Operating Manual 1411.4060.32 ─ 09
Example:
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
'1.2.3.4'
sets the IP address of the default gateway.
Manual operation: See "Default Gateway" on page 93
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK <Mask>
Sets the subnet mask.
Parameters:
<Mask> string
Example:
SYSTem:COMMunicate:NETWork:IPADdress:SUBNet:
MASK '255.255.0.0'
determines the subnet mask.
Manual operation: See "Subnet Mask" on page 93
:SYSTem:COMMunicate:NETWork:MACaddress <MacAddress>
Queries the MAC address of the network adapter.
Parameters:
<MacAddress> string
Example:
SYST:COMM:NETW:MAC
queries the MAC address.
Manual operation: See "MAC Address" on page 93
:SYSTem:COMMunicate:NETWork:STATus?
Queries the network configuration state.
Return values:
<State> 0 | 1 | OFF | ON
Usage: Query only
Manual operation: See "Network Status" on page 91
:SYSTem:COMMunicate:NETWork:RESTart
Restarts the network connection to the instrument, terminates the connection and sets
it up again.
Example:
SYSTem:COMMunicate:NETWork:RESTart
Usage: Event
Manual operation: See "Restart Network" on page 93
SYSTem Subsystem