Remote Control Commands
R&S
®
SMC100A
271Operating Manual 1411.4060.32 ─ 09
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority <Priority>
The command selects which power measurement result (average or peak power) is
indicated when permanent display is active.
Parameters:
<Priority> AVERage | PEAK
*RST: PEAK
Example:
SENS1:DISP:PERM:STAT ON
the permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
the measured average power is indicated.
Manual operation: See " Display Priority " on page 151
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
The command queries the current filter length for auto filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE AUTO)
Return values:
<Auto> float
Range: 1 to 65536
Example:
SENS1:FILT:TYPE AUTO
selects auto filter mode for the power sensor connected to the
SENSOR connector.
SENS1:FILT:LENG:AUTO?
queries the automatically set filter length.
Response: 1024
Usage: Query only
Manual operation: See "Filter Length" on page 154
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
The command selects the filter length for user filter mode (SENSe:POWer:FILTer:
TYPE USER). As the filter length works as a multiplier for the time window, a constant
filter length results in a constant measurement time. Values 1 and 2^n are settable.
The time window is fixed to 20 ms.
Parameters:
<User> float
Range: 1 to 65536
*RST: 1
SENSe, READ and INITiate Subsystems