Instrument Function
R&S
®
SMC100A
151Operating Manual 1411.4060.32 ─ 09
To query the availability of a sensor at a given connector, use the command
SENSe<ch>[:POWer]:STATus[:DEVice]? on page 275.
Remote command:
:INITiate<ch>[:POWer]:CONTinuous on page 269
Level (Peak)
With certain power sensors only, for example R&S NRP-Z81.
Indicates the measured peak level value with the selected unit.
Remote command:
:READ<ch>[:POWer]? on page 269
Level (Avg.)
Indicates the measured level value with the selected unit.
Remote command:
:READ<ch>[:POWer]? on page 269
Unit
Selects the unit used for result display.
The power sensor provides the measured value in Watt.
In which unit the measured value is indicated is selected here and might be Watt, dBm
or dBuV.
Remote command:
:SENSe<ch>:UNIT[:POWer] on page 277
Permanent Display State
Activates the permanent indication of the power measurement result in the upper right
corner of the block diagram. The instrument shows the type of sensor, the correspond-
ing connector, the measurement source and - if set - the offset.
It is possible to switch the permanent display active for several sensors. In this case,
the instrument indicates the values of the sensor with the lowest port number in the
display.
Remote command:
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe on page 270
Display Priority
Determines whether the instrument displays the measured average or the peak power
permanently on the screen.
To select the peak power display, it is required that the R&S NRP-Zxx sensor supports
this feature. On power-on or connecting a sensor the average power value is set by
default.
To enable the permanent display in the block diagram, select Permanent Display
State .
RF Block