Remote Control Commands
R&S
®
SMC100A
274Operating Manual 1411.4060.32 ─ 09
:SENSe<ch>[:POWer]:LOGGing:STATe <State>
Activates the recording of the power values, measured by a connected R&S NRP-Z
power sensor.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENS:LOGG:STAT ON
activates recording of the power measurement of the first sen-
sor.
Manual operation: See "Enable Logging" on page 155
:SENSe<ch>[:POWer]:FREQuency <Frequency>
The command sets the RF frequency of the source if the user source is selected
(SENSe[:POWer]:SOURce USER).
Parameters:
<Frequency> float
*RST: 1 GHz
Example:
SENS:SOUR USER
selects user-defined source.
SENS:FREQ 2.44 GHz
enters the RF frequency of the source which is 2.44 GHz.
Manual operation: See " Frequency " on page 152
:SENSe<ch>[:POWer]:OFFSet <Offset>
The command enters a level offset which is added to the measured level value after
activation with command SENSe[:POWer]:OFFSet:STATe ON. This allows e.g. an
attenuator in the signal path to be considered.
Parameters:
<Offset> float
Range: -100.0 to 100.0
*RST: 0
Default unit: dB
Example:
SENS:POW:OFFS 10.0
sets a level offset of 10 dB
Manual operation: See " Level Offset " on page 153
:SENSe<ch>[:POWer]:OFFSet:STATe <State>
The command activates the addition of the level offset to the measured value. The
level offset value is set with command SENSe[:POWer]:OFFSet.