Remote Control Commands
R&S
®
SMC100A
249Operating Manual 1411.4060.32 ─ 09
Return values:
<Data> block data
Example:
HCOP:DEV:LANG JPG
HCOP:DATA?
Transfers the hardcopy to the remote client.
Usage: Query only
:HCOPy:DEVice <Device>
Defines the output device. The hardcopy can be output in a file. The HCOPy:FILE:…
commands are used for configuration. The file is accessible using the MMEM:.. com-
mands. In addition, the hardcopy data can be directly transferred to the remote client
using command :HCOPy:DATA?.
Parameters:
<Device> FILE
*RST: FILE
Example:
HCOP:DEV FILE
The hardcopy is stored in a file.
Manual operation: See "Destination" on page 107
:HCOPy:DEVice:LANGuage <Language>
Selects the bitmap graphic format for the screenshot.
You can also retrieve the data using command HCOP:DATA?. This command is an
alias to command HCOPy:IMAGe:FORMat.
Parameters:
<Language> BMP | JPG | XPM | PNG
*RST: PNG
Example:
HCOP:DEV:LANG BMP
Selects bitmap as image format.
HCOP:FILE '/usb/HCopy'
Defines the directory, path and filename for storing the hard-
copy.
HCOP
Triggers the hardcopy generation.
Manual operation: See " File Options " on page 108
:HCOPy[:EXECute]
Triggers the generation of a hardcopy.
Example:
HCOP
Triggers the generation of a hardcopy of the current display.
HCOPy Subsystem