Remote Control Commands
R&S
®
SMC100A
321Operating Manual 1411.4060.32 ─ 09
Parameters:
<PowCntrlTarget> float
Range: -50 to 30
Increment: 0.01
*RST: -10
Example:
SENS:UNIT dBm
selects unit dBm for setting the target level value.
POW:SPC:TARG -10
sets -10 dBm target level.
Manual operation: See "Target Level" on page 139
[:SOURce<hw>]:POWer:STARt
<Start>
Sets the start level for the RF sweep.
Note: You can select any level within the setting range. The range is defined by this
start value and the [:SOURce<hw>]:POWer:STOP value.
A defined offset ([:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet)
affects the level values according to the formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
<Start> float
Determines the first level value of the sweep setting range.
Range: full specified level range
Increment: see the data sheet: Level sweep > Step size setting
resolution
*RST: -30
Example:
POW:STAR -20 dBm
sets the start level for the level sweep to -15 dBm for RF output
A.
Manual operation: See " Start Level - Level Sweep " on page 167
[:SOURce<hw>]:POWer:STEP[:INCRement] <Increment>
Sets the step width for POW:STEP:MODE USER.
To adjust the level step by step with this step size, use the POW:UP and POW:UP com-
mands.
Note: This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the level accordingly, when you work in user-defined step
mode.
SOURce Subsystem