Remote Control Commands
R&S
®
SMC100A
320Operating Manual 1411.4060.32 ─ 09
Example:
:POW:SPC:DEL 2 ms
the sensor starts the next reading 2 ms after the level adjust-
ment.
Manual operation: See "Delay Time" on page 139
[:SOURce<hw>]:POWer:SPC:PEAK <PowCntrlPeak>
Activates power control by means of the peak power values, provided the power sen-
sor supports this function.
Parameters:
<PowCntrlPeak> 0 | 1 | OFF | ON
*RST: 0
Example:
POW:SPC:PEAK ON
uses the measured peak power for power control.
Manual operation: See "Use Peak Power" on page 139
[:SOURce<hw>]:POWer:SPC:SELect <PowCntrlSelect>
Defines the currently selected sensor to be used for power control.
Parameters:
<PowCntrlSelect> SENS1 | SENS2 | SENS3 | SENS4
*RST: SENS1
Example:
POW:SPC:SEL SENS2
selects the sensor connected to a second USB interface for
power control.
Manual operation: See " Sensor " on page 138
[:SOURce<hw>]:POWer:SPC:STATe <PowCntrlState>
Activates power control using the selected sensor. The control loop periodically adjusts
the generator output. After switching off, the running loop is completed.
Parameters:
<PowCntrlState> 0 | 1 | OFF | ON
*RST: 0
Example:
POW:SPC:STAT ON
activates power control.
Manual operation: See "State" on page 138
[:SOURce<hw>]:POWer:SPC:TARGet <PowCntrlTarget>
Sets the nominal level expected at the input of the sensor. To define the unit of the
power value, use command :SENSe<ch>:UNIT[:POWer] on page 277.