Remote Control Commands
R&S
®
SMC100A
319Operating Manual 1411.4060.32 ─ 09
The level entered with this command corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note: The SCPI command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude] sets the level of the "Level" display, i.e. the level containing offset.
Parameters:
<Power> Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The values are given in the data sheet.
Increment: 0.01
*RST: -30
Example:
SOUR:POW:POW 15
sets the RF level at output to 15 dBm.
Manual operation: See "Amplitude" on page 131
[:SOURce<hw>]:POWer:SPC:CRANge <PowCntrlCRange>
Defines the capture range of the power control system.
Within the range:
Target Level +/- Catch Range
the power control locks and tries to achieve the target level. Readings outside the
range are not considered.
Parameters:
<PowCntrlCRange> float
Range: 0 to 50
Increment: 0.01
*RST: 30
Default unit: dB
Example:
POW:SPC:CRAN 15
sets the capture range to +/- 15 dB.
Manual operation: See "Catch Range +/-" on page 139
[:SOURce<hw>]:POWer:SPC:DELay <PowCntrlDelay>
Defines a waiting period between the level adjustment of the generator and the next
measurement of the power sensor.
Parameters:
<PowCntrlDelay> integer
Range: 0 to 1000
*RST: 0
SOURce Subsystem