Instrument Function
R&S
®
SMC100A
167Operating Manual 1411.4060.32 ─ 09
Remote command:
[:SOURce<hw>]:SWEep:POWer:EXECute on page 339
:TRIGger<hw>:PSWeep[:IMMediate] on page 367
:TRIGger<hw>[:SWEep][:IMMediate] on page 368
Start Level - Level Sweep
Sets the start level.
Remote command:
[:SOURce<hw>]:POWer:STARt on page 321
Stop Level - Level Sweep
Sets the stop level.
Remote command:
[:SOURce<hw>]:POWer:STOP on page 322
Current Level - Level Sweep
Displays the current level.
If "Step" is set, the level for the next level step of the sweep is entered here.
Remote command:
[:SOURce<hw>]:POWer:MANual on page 317
Shape - RF Level Sweep
Selects the cycle mode for a sweep sequence (shape).
"Sawtooth"
One sweep runs from the start level to the stop level. The subsequent
sweep starts at the start level again, i.e. the shape of sweep
sequence resembles a sawtooth.
"Triangle"
One sweep runs from start to stop level and back, i.e. the shape of
the sweep resembles a triangle. Each subsequent sweep starts at the
start level again.
Remote command:
[:SOURce<hw>]:SWEep:POWer:SHAPe on page 341
Retrace - RF Level Sweep
Activates that the signal changes to the start level value while it is waiting for the next
trigger event. It allows you to shift down the power during the waiting period.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single", see Mode - Level Sweep .
RF Block