Instrument Function
R&S
®
SMC100A
102Operating Manual 1411.4060.32 ─ 09
Enable LAN Interface ← LAN Services
Enables the LAN interface in general, and thus provides remote access via all
unlocked services.
Note: The activated LAN services will not be assigned until you enter the Security
Password and confirm with Accept.
Enable LAN Services individually ← LAN Services
Enables or disables the following interface services individually.
"SCPI over LAN"
activates access over LAN to remotely control the instrument using
SCPI (Standard Commands for Programmable Instruments) com-
mands.
"VNC"
activates access via VNC (Virtual Network Computing) interface, a
graphical desktop sharing system that uses RFB protocol to remotely
control the instrument.
"SSH"
activates access via SSH (Secure Shell), a network protocol for
secure data communication.
"HTTP"
activates access via HTTP (Hyper Text Transfer Protocol), the appli-
cation protocol for hypermedia information systems.
"FTP"
activates access via FTP (File Transfer Protocol), used to transfer
files from a host to the instrument and vice versa.
"SMB (Samba)"
activates access to SMB (Server Message Block), used for providing
shared access to files, printers and serial ports of a network.
"Avahi (Zeroconf)"
activates Avahi, a service for automatic configuration of the instru-
ment in a network environment.
"Software Update"
allows updating the instrument firmware via the LAN interface.
For more information on this topic see the release notes of the instru-
ment, provided on the Internet at the download site or the Rohde &
Schwarz Signal Generator home page.