Remote Control Basics
R&S
®
SMC100A
205Operating Manual 1411.4060.32 ─ 09
2. In the R&S VISA "Basics" tab, enter a SCPI command, e.g. "*IDN?" and confirm
with "Query".
The instrument is switched to remote control when it receives a command from the
controller.
3. Select "Read" to obtain the instrument response.
Tip: If the "Show Log" checkbox is checked R&S VISA displays each VISA function
call in the log-view on the left. If you check the "Write Log" checkbox the log-view
entry is written to the log file as well. You can operate the log-view in two modes:
the "Live Mode" shows only the most recent messages whereas the "View Mode"
allows you to scroll the history.
4. To set, e.g. the frequency, enter SOUR1:FREQ 4 GHz and select "Write".
To check the performed setting, SOUR1:FREQ? and select "Read".
The instrument response is 4000000000, i.e. the frequency is returned in Hz.
While remote control is active, the "Remote" icon in the status bar indicates that the
instrument is in remote control mode. The operation via the front panel or via
mouse and keyboard are locked, allowing a remote control program to be per-
formed without interruption.
On the display, keys and entry fields are grayed out and cannot be activated or
modified, but you can still open dialogs, for example to verify settings.
5. To disable the access to the dialogs, use the command SYST:KLOC ON.
6. To prevent unintentional return to manual operation, use the command &LLO.
See also Chapter 6.1.3.4, "LAN Interface Messages", on page 195.
The instrument switches to "Remote LLO" state. The LOCAL key is disabled.
Starting a Remote Control Session