Remote Control Commands
R&S
®
SMC100A
331Operating Manual 1411.4060.32 ─ 09
Manual operation: See "Deactivate RF Output (if external reference is missing)"
on page 128
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue <Value>
Specifies the frequency correction value (adjustment value).
Parameters:
<Value> integer
Range: 0 to maximum value (see data sheet)
Increment: see data sheet
*RST: ---
Example:
ROSC:ADJ:VAL 456
Sets the adjustment value to 456.
Manual operation: See "Adjustment DAC Value" on page 128
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] <State>
Determines whether the calibrated (OFF) or a user-defined (ON) adjustment value is
used for fine adjustment of the frequency.
If user-defined values are used, the instrument is no longer in the calibrated state.
However, the calibration value is not changed and the instrument resumes the calibra-
ted state after sending the command :SOURce:ROSCillator:INTernal:ADJust:
STATe OFF.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: -
Example:
ROSC:SOUR INT
Selects the internal source.
ROSC:ADJ ON
Activates use of a user-defined adjustment value.
ROSC:ADJ:VAL 1400
Sets the adjustment value to 1400.
Manual operation: See "Adjustment Active" on page 128
[:SOURce]:ROSCillator:SOURce <Source>
Selects the reference frequency source.
SOURce Subsystem