The command is only effective for sweep mode "Single" (SOUR:SWE:FREQ:MODE
AUTO in combination with TRIG:FSW:SOUR SING).
The command corresponds to the manual control "Execute Single Sweep".
Example:
SWE:FREQ:MODE AUTO
sets the triggered sweep mode, i.e. a trigger is required to start
the sweep.
TRIG:FSW:SOUR SING
sets the "Single" trigger mode, i.e. a trigger starts a single
sweep.
TRIG:FSW
starts a single RF frequency sweep.
Usage: Event
Manual operation: See " Execute Single Sweep - Frequency Sweep " on page 160
:TRIGger<hw>:LFFSweep
Usage:
Event
Manual operation: See " Execute Single Sweep " on page 183
Immediately starts an LF frequency sweep.
The command is effective in sweep mode "Single" (LFO:SWE:MODE AUTO in combina-
tion with TRIG:LFFS:SOUR SING).
:TRIGger<hw>:LFFSweep:SOURce <Source>
Sets the trigger source for the LF sweep. The trigger is triggered by the com-
mand :SOURce:LFOutput:SWEep[:FREQuency]EXECute.
The names of the parameters correspond directly to the various settings under manual
control. SCPI uses other names for the parameters; these names are also accepted by
the instrument. The SCPI names should be used if compatibility is an important consid-
eration.
An overview of the various names is given in the Table 7-2.