Risk of causing instrument unusability
The instrument is equipped with the Linux operating system. Additional software can
therefore be installed on the instrument. The use and installation of additional software
may impair instrument function. Thus, run only programs that Rohde & Schwarz has
tested for compatibility with the instrument software.
The drivers and programs used on the instrument under Linux have been adapted to
the instrument. Existing instrument software must always be modified using only
update software released by Rohde & Schwarz.
The configuration of the operating system is optimally adapted to signal generator
functions in the factory. Changes in the system setup are only required when peripher-
als like keyboard are installed or if the network configuration does not comply with the
default settings (see Chapter 2.6.1, "Connecting the Instrument to the Network",
on page 34). After the R&S SMC is started, the operating system boots and the
instrument firmware is started automatically.
Accessing Operating System
No access to the operating system is required for normal operation.
All necessary system settings can be made in the "Setup" dialog.
The R&S SMC provides an internal flash memory, it does not contain a disk drive. The
flash memory holds the operating system, the firmware and the stored data. Data
transfer is possible via a memory stick connected to a USB interface. The memory
stick and the flash memory are accessed via the "File Manager".
Accessing the File System
The instrument also supports two standard methods to access the file system form a
remote client:
●
FTP (file transfer protocol)
●
File sharing according to the SAMBA/SMB (server message block) protocol.
Both methods allow the access to the folder /var/user/share.
Default password
The FTP and SAMBA/SMB file access use the user "instrument" with the default pass-
word "instrument".
It is highly recommenced that you change this password in the Chapter 5.2.3.15,
"Security", on page 99 dialog before connecting the instrument to the network!
Refer also to Application Note 1GP72 "Connectivity of Signal Generators", provided on
the Rohde & Schwarz website, at http://www.rohde-schwarz.com/appnotes.