SME LIST Mode
1038.6002.02 2.145 E-13
TRIGGER input
BLANK-output
MARKER-output
Frequency
Fig. 2-58 Signal example LIST mode: MODE = EXT-STEP
The LIST menu offers access to settings for the LIST mode.
Menu selection: LIST
FREQUENCY
LEVEL
MODULATION
DIGITAL MOD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
HELP
LEVEL
- - -.-
- - -.- - -
MODE
OFF AUTO SINGLE STEP EXT-SINGLE EXT-STEP
RESET LIST
DWELL 10.0 ms
CURRENT INDEX 1
LEARN
SELECT LIST... CURRENT: LIST2
DELETE LIST...
FUNCTION FILL INSERT DELETE EDIT/VIEW
FREQ
LIST
EXECUTE SINGLE LIST
Fig. 2-59 Menu LIST - OPERATION page
MODE
Selection of the operating mode (cf. Section 2.9.1, operating modes).
IEC bus short commands : :SOUR:FREQ:MODE LIST;
:SOUR:LIST:MODE AUTO;
:TRIG:LIST:SOUR SING
EXECUTE SINGLE LIST Ð
Starts a single run of a list. This menu option is only visible if MODE
SINGLE is selected.
IEC bus short command : :TRIG:LIST