DIAGnostic SME
1038.6002.02 3.22 E-13
:DIAGnostic:INFO:CCOunt:POWer?
The command queries the number of switch-on processes.
Example: :DIAG:INFO:CCO:POW? Response: 258
:DIAGnostic:INFO:MODules?
The command queries the modules existing in the instrument with their model and
state-of-modification numbers. The response supplied is a list in which the different entries are
separated by commas. The length of the list is variable and depends on the equipment of the
instrument. Each entry consists of three parts which are separated by means of blanks:
1. Name of module; 2. Variant of module in the form VarXX (XX = 2 digits)
3. Revision of module in the form RevXX (XX = 2 digits)
Example :DIAG:INFO:MOD? Response: FRO Var01 Rev00, DSYN Var03 Rev12, to
:DIAGnostic:INFO:OTIMe?
The command reads out the internal operating-hours counter. The response supplies the number
of hours the instrument has been in operation up to now.
Example: :DIAG:INFO:OTIM? Response: 19
:DIAGnostic:INFO:SDATe?
The command queries the date of software creation. The response is returned in the form year,
month, day.
Example: :DIAG:INFO:SDAT? Response: 1992, 12, 19
:DIAGnostic:[:MEASure]
The commands which trigger a measurement in the instrument and return the measured value
are under this node.
:DIAGnostic[:MEASure]:POINt?
The command triggers a measurement at a measuring point and returns the voltage measured.
The measuring point is specified by a numeric suffix (cf. service manual, stock no.
1039.1856.24).
Example: :DIAG:MEAS:POIN2? Response: 3.52
:DIAGnostic:XMEM:CHECksum
The command for calculating and querying the checksum are provided in this node. Contrary to
the display in local mode, the values are returned as decimal numbers.
:DIAGnostic:XMEM:CHECksum:CALCulate
This command triggers the calculation of all four checksums. The results depend on the start and
stop addresses selected and on the operating mode of the memory extension. This command is
recommended whenever a checksum is to be read out and when setups or memory extension
data have been modified since the last calculation.
Example: :DIAG:XMEM:CHEC:CALC Action without query