34 Operating Instructions Cubis
®
MCE
Installation
5.5 Preparing Below-cell Weighing
Samples that cannot be placed on the weighing pan can be suspended
below the device base and weighed using the below-cell weighing
equipment.
In legal metrology:
Do not use the below-cell weighing equipment.
Do not open the cover of the below-cell weighing equipment.
Material: 1 soft support base
Requirements
The weighing pan and the associated components have not been set up.
For a device with an analytical draft shield or flat glass draft shield: The
side panels and upper panel have not been fitted.
Procedure
t Turn the device on its side and place it on the soft support base.
t Remove the hook for below-balance weighing (1) from the retainer on
the underside of the base of the device.
t Pull the cover of the below-balance weighing equipment (1) out.
1
1