Operating Instructions Cubis
®
MCE 65
Operation
8.10 Printing Results
The results of the calibration can be printed in a GLP printout.
Procedure
t Set the GLP printout in the menu (setting parameters, see
Chapter “Parameters in the “Settings for Printouts” Submenu”,
page 53):
t Calibrate the device.
y Once the calibration is complete:
y The printing process starts.
y The GLP footer is printed.
8.11 Weighing and Printing with ID Marking
The device and every lot can be assigned an ID number. The ID numbers are
provided in the GLP header during GLP-compliant printing.
Requirements
The ID marking is entered in the menu, e.g. “BATCH1” (see Chapter “7.3.5
Parameters in the “Input” Main Menu”, page 56).
The printing of the lot ID line in the GLP printout is activated in the menu
(see Chapter “7.3.5 Parameters in the “Input” Main Menu”, page 56).
The GLP-compliant printout is activated (see Chapter “7.3.5 Parameters
in the “Input” Main Menu”, page 56).
The format for the printout is set to 22 characters per line (configuration
parameters, see Chapter “7.3.3 Parameters in the “Data Output” Main
Menu”, page 52).
Procedure
t Start the printout. To do so, press the [Print] key.
t The GLP header is printed with the ID marking set in the menu and the
current weight value.
t The [GLP] button appears in the operating display.
t To exit the GLP printout: Press the [GLP] button.
y The GLP footer is printed.
----------------------
23-Apr-2020 13:06
Sartorius
Mod. MCE5202S
Ser. no. 12345678
Ver. 09-05-00.09.01
ID CHARGE01
----------------------
L ID
G + 50.12 g
----------------------
23-Apr-2020 13:07