64 Operating Instructions Cubis
®
MCE
Operation
t Place the indicated calibration weight on the balance. The device is
automatically adjusted after calibration, if the calibration weight placed
on the device is within the specified limits.
y The [+] sign is shown if the applied weight is too high.
y The [-] sign is shown if the applied weight is too low.
y The [CAL.END] message indicates the end of calibration.
t Remove the calibration weight after calibration is complete.
Manually Starting Adjustment after Calibration
Requirements
− The weighing pan is unloaded.
− The operating display shows a stable weight value.
− The device is not locked against external calibration.
Procedure
t To set the external calibration in the menu: In the “SETUP” / “BALANCE””
menu, for the “CAL./ADJ.” parameter, select external calibration with
preset calibration weight or external calibration with user-defined
calibration weight as described above.
t To set the sequence of the adjustment function to calibration without
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu,
for the “CAL.SEQ.” parameter, select the “CAL/ADJUST” configuration
value.
t Exit the menu.
t Press the [Zero] key.
t Press the [Adjust] button.
t If automatic leveling is set for a model with motorized adjusting feet: The
device levels itself automatically.
t Place the indicated calibration weight on the balance. The calibration
can only be started, if the calibration weight used is within the defined
limits.
y The [+] sign is shown if the applied weight is too high.
y The [-] sign is shown if the applied weight is too low.
t Press the [Confirm] button on the keypad. This starts the calibration.
y The [CAL.END] display indicates the end of calibration.
t Remove the calibration weight after calibration is complete.