Operating Instructions Cubis
®
MCE 59
Operation
8.3 Opening and Closing the Motorized Draft
Shield (Only for Devices with a Motorized
Draft Shield)
Pressing the [Change] key enables the motorized side and upper draft
shield panels to be opened and closed. The [Change] key can be used to
control up to three doors simultaneously.
Procedure
t Switch on the device.
t Close all draft shield doors.
t To determine how far a door is opened by pressing the [Change] key:
Manually push the door open to the desired position.
t If several doors are to be controlled by the [Change] key simultaneously:
Manually push the desired doors open to the desired position.
t Press the [Change] key.
y The opened door is closed.
y If several doors are open: All open doors are closed.
y When the [Change] key is next pressed, the door is opened or closed.
8.4 Preparing Weighings
The device must be prepared before every weighing.
Requirements
The device is ready for use
NOTICE
Chemicals may damage the device or accessories!
Chemicals can attack the device or the connected accessories internally
and externally. This may damage the device and accessories.
t Use appropriate containers when weighing chemicals.
Procedure
t Level the device.
t Zero the device for calibration. To do so, press the [Zero] key.
t If the device cannot be zeroed: Remove the sample to be weighed and
re-zero the device.
t Calibrate the device.