5.7 Installing a Device with a Frame Draft Shield
5.7.1 Positioning the Weighing Pan and Associated
Components
Procedure
t Insert the pin on the pan support into the clip on the pan retainer.
t Push the pan support down onto the pan retainer until the pan support
lies parallel to the device housing.
t Place the shield plate (3) on the device housing.
t Place the weighing pan (2) on the pan support.
t Place the frame draft shield (1) on the shield plate (3).
5.8 Connecting the Electronics Module
(Onlyfor Semi-microbalance)
Procedure
t Connect the connection cable to the electronics module’s weighing
module connection.
t Connect the other end of the connection cable to the weighing
module’s electronics module connection.
t To lock the connection cable: Lock the plugs of the connection cable
onto both connections with two clicks in each case.
t NOTICE Damage to the device due to incorrect connection!
t Check the correct fit of the plug contacts.
t There should be no tension on the connection cable, e.g. do not
install directly against a wall.
1
2
3