Operating Instructions Cubis
®
MCE 63
Operation
t To start adjustment: Press the [Confirm] button on the keypad.
y The device is calibrated and the internal calibration weight is removed.
y The [CAL.END] display indicates the end of calibration.
8.9 Externally Calibrating and Adjusting the
Device (Not for Conformity-assessed
Models)
External Calibration with Factory Set Weight Value
Procedure
t In the “SETUP” / “BALANCE” menu, for the “CAL./ADJ.” parameter,
select the “EXT.CAL.” configuration value.
Set the User-defined Weight Value
A weight value can be set for the calibration weight. This user-defined
weight value is used for the external calibration.
If no user-defined weight value is set: The preset weight value is used.
Procedure
t In the “INPUT” menu, select the “CAL.WT.” configuration value.
t Enter and confirm the desired weight value.
t To use the user-defined weight value for the next calibration: In the
“Balance” menu, for the “CAL./ADJ.” parameter, select the “E.CAL.USR.”
configuration value.
Calibration and Subsequent Adjustment
Requirements
− The weighing pan is unloaded.
− The operating display shows a stable weight value.
− The device is not locked against external calibration.
− Keep calibration weight ready for use.
Procedure
t To set the external calibration in the menu: In the “SETUP” / “BALANCE”
menu, for the “CAL./ADJ.” parameter, select external calibration with
preset calibration weight or external calibration with user-defined
calibration weight as described above.
t To set the sequence of the adjustment function to calibration and
subsequent automatic adjustment: In the “SETUP” / “BALANCE” menu,
for the “CAL.SEQ.” parameter, select the “ADJUST” configuration value.
t Exit the menu.
t Press the [Zero] key.
t Press the [Adjust] button.
t If automatic leveling is set for a model with motorized adjusting feet:
Thedevice levels itself automatically.