EasyManua.ls Logo

Sartorius Cubis MCE - Models MCE623 S | MCE623 P | MCE323 S | MCE5202 S

Sartorius Cubis MCE
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions Cubis
®
MCE 89
Technical Data
15.7.5 Models MCE623S | MCE623P | MCE323S | MCE5202S
MCE623S MCE623P MCE323S MCE5202S
Unit Value Value Value Value
Scale interval (d) mg 1 1 | 2 | 5 1 10
Maximum capacity (Max) g 620 150 | 300 | 620 320 5200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 0.7 1 0.7 6
Standard deviation of the load values,
typical value
mg 0.4 0.4 0.4 2
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 0.7 1 0.7 6
Standard deviation of the load values,
typical value
mg 0.5 0.5 0.5 2
Linearity deviation
Tolerance mg 2 5 2 10
Typical value mg 0.6 1.5 0.6 5
Deviation when load is off-center, positions
according to OIML R76
Test weight g 200 200 200 2000
Tolerance mg 2 4 2 10
Typical value mg 1 3 1 5
Sensitivity drift between +10°C and +30°C ppm/K 2 2 2 2
Tare-maximum capacity: Less than 100 %
of maximum capacity
Accuracy class according to Directive
2014/31/EU
II II II I
Verification scale interval (e) according to
Directive 2014/31/EU
mg 10 10 10 100
Minimum load (Min) according to Directive
2014/31/EU
mg 20 20 20 1000
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 820 820 820 8200
Typical minimum initial weighing mg 820 820 820 8200
Typical stabilization time s 0.8 0.8 0.8 0.8
Typical measurement time s 1 1 1 1

Table of Contents